Calibration Program

Document Control

Field Value
Document ID FS-CAL-001
Version 2.1
Effective Date 2026-01-05
Owner Maintenance Manager
Approver Food Safety Team Leader
Review Frequency Annual

1. Calibration Policy

1.1 Purpose

To ensure all measuring, monitoring, and test equipment used for food safety control, quality verification, and process control is calibrated at defined intervals against traceable standards, ensuring accuracy and reliability of results.

1.2 Scope

All equipment used in the measurement of parameters critical to food safety and quality, including: - Temperature (ovens, chillers, freezers, heat exchangers, water heaters) - Weight / mass (processing scales, checkweighers, lab balances) - Pressure (CIP pumps, oil presses, steam systems) - Flow rate (CIP flow meters, ingredient dosing) - pH (CIP return, water treatment) - Conductivity / TDS (CIP final rinse, water treatment) - Humidity (storage areas) - Metal detection / X-ray (reject verification) - Time (process timers, cook times) - Analytical instruments (ELISA readers, ATP luminometers) - Chemical concentration (dispensers, test kits)

1.3 Key Principles

  1. Traceability: All calibrations must be traceable to national or international standards (NIST, UKAS, ISO 17025 accredited laboratories).
  2. Frequency: Defined per equipment type (Section 2). Adjust frequency based on drift trends.
  3. Before-use checks: Operators perform daily verification on critical devices before production start.
  4. Records: All calibration records retained for minimum 5 years.
  5. Only calibrated equipment may be used for food safety / quality decisions.
  6. Out-of-tolerance results trigger formal investigation (Section 4).

1.4 Roles & Responsibilities

Role Responsibility
Maintenance Manager Program owner; schedule execution; external calibration vendor management
QA Manager Approval of calibration tolerances; review of out-of-tolerance investigations
Production Operators Daily pre-use checks (e.g., checkweigher, metal detector verification)
External Calibration Lab Perform annual calibrations with ISO 17025 accredited scope

2. Calibration Equipment Register

2.1 Master Register

ID Equipment Location Range / Capacity Required Accuracy Standard Used Frequency Internal/ External Procedure Ref
CAL-T-001 Oven (Hot Air Dryer) — D-301 Drying Room 30–180 °C ±1.0 °C Fluke 1523 Reference Thermometer 6 months Internal CAL-PROC-01
CAL-T-002 Oil Press Heater — H-301 Oil Press Area 20–120 °C ±0.5 °C Fluke 1523 Reference Thermometer Monthly (internal) + annual (external) Both CAL-PROC-01
CAL-T-003 Cold Storage — CS-01 Warehouse -5 to 10 °C ±1.0 °C Pt100 probe + calibrated logger Monthly (internal) + annual (external) Both CAL-PROC-01
CAL-T-004 Blast Chiller — BC-01 Packaging Area -20 to 5 °C ±1.0 °C Pt100 probe Quarterly Internal CAL-PROC-01
CAL-T-005 Steam Boiler Temp Sensor — B-101 Boiler Room 100–200 °C ±2.0 °C Thermocouple calibrator Annual External CAL-PROC-02
CAL-W-001 Platform Scale (500 kg) — S-101 Receiving Bay 0–500 kg ±0.1 kg NIST Class F weights (10 × 20 kg) Monthly (internal) + annual (external) Both CAL-PROC-03
CAL-W-002 Bench Scale (10 kg) — S-102 Production Floor 0–10 kg ±1 g NIST Class F weights (1 kg + 5 kg) Weekly (internal) + annual (external) Both CAL-PROC-03
CAL-W-003 Analytical Balance (220 g) — LB-101 QA Lab 0–220 g ±0.1 mg NIST Class 1 weights (100 g, 50 g, 20 g) Quarterly (internal) + annual (external) Both CAL-PROC-04
CAL-W-004 Checkweigher — CW-301 Packing Line 0–30 kg ±5 g NIST Class F weight set Daily (auto check) + monthly (internal) + annual (external) Both CAL-PROC-03
CAL-P-001 CIP Supply Pump Pressure Gauge — P-301 CIP Skid 0–10 bar ±0.1 bar Druck DPI 611 Pressure Calibrator 6 months Internal CAL-PROC-05
CAL-P-002 Oil Press Hydraulic Gauge — HY-101 Oil Press 0–400 bar ±2 bar Druck DPI 611 Pressure Calibrator 6 months Internal CAL-PROC-05
CAL-P-003 Steam Pressure Gauge — ST-101 Boiler Room 0–16 bar ±0.2 bar Druck DPI 611 Pressure Calibrator Annual External CAL-PROC-06
CAL-F-001 CIP Flow Meter — FM-301 CIP Return Line 0–20 m³/h ±0.5% of reading Prover / master meter (tank test) Annual External CAL-PROC-07
CAL-F-002 Water Flow Meter — FM-201 Water Inlet 0–50 m³/h ±1.0% Master meter comparison Biennial External CAL-PROC-07
CAL-pH-001 pH Meter — pH-101 QA Lab 0–14 pH ±0.02 pH pH buffer standards (pH 4.0, 7.0, 10.0) Daily (verification) + monthly (full cal) Internal CAL-PROC-08
CAL-pH-002 Inline pH Probe — pH-201 CIP Return 0–14 pH ±0.1 pH pH buffer loop + lab verification Weekly (verification) + 6 monthly (full cal) Both CAL-PROC-08
CAL-CD-001 Conductivity Meter — CD-101 QA Lab 0–2000 µS/cm ±1% of reading KCl conductivity standard (1413 µS/cm) Monthly Internal CAL-PROC-09
CAL-CD-002 Inline Conductivity Probe — CD-201 CIP Return 0–2000 µS/cm ±2% of reading KCl standard loop check Monthly (verification) + annual (full cal) Both CAL-PROC-09
CAL-MD-001 Metal Detector — MD-301 Packing Line Ferrous: 0.5–5.0 mm; Non-ferrous: 0.8–5.0 mm; SS: 1.2–5.0 mm Detects Fe 1.0 mm, NF 1.5 mm, SS 2.0 mm Calibration test sticks (Fe, NF, SS) Hourly (run test sticks) + monthly (internal cal) + annual (external) Both CAL-PROC-10
CAL-XR-001 X-ray Inspection — XR-301 Packing Line Resolution 0.3 mm Detects SS 0.5 mm Calibration test piece (0.5 mm, 1.0 mm SS spheres) Hourly (run test piece) + monthly (internal) + annual (external) Both CAL-PROC-11
CAL-AT-001 ATP Luminometer — LM-101 QA Lab 0–9999 RLU ±5% at 1000 RLU ATP calibration standard (provided by manufacturer) Annual (return to manufacturer) External CAL-PROC-12
CAL-AT-002 ATP Luminometer — LM-102 Production Floor 0–9999 RLU ±5% at 1000 RLU ATP calibration standard Annual (return to manufacturer) External CAL-PROC-12
CAL-EL-001 ELISA Plate Reader — EPR-101 QA Lab 450 nm, 620/650 nm absorbance ±0.005 OD Neutral density filters (0.5, 1.0, 1.5 OD) Annual External CAL-PROC-13
CAL-TI-001 Process Timer — TIM-101 Production Line 0–99 min ±1 sec Stopwatch (traceable) Quarterly Internal CAL-PROC-14
CAL-TI-002 Oven Timer — TIM-201 Drying Room 0–99 min ±5 sec Stopwatch (traceable) Quarterly Internal CAL-PROC-14
CAL-HU-001 Humidity Logger — HL-101 Cold Storage 20–90% RH ±3% RH Humidity generator / saturated salt standards Annual External CAL-PROC-15

2.2 Colour Coding / Label System

Each calibrated instrument displays a calibration label with:

     ┌──────────────────────────┐
     │   CALIBRATION STATUS     │
     │  EQUIPMENT ID: CAL-T-002 │
     │  CAL DATE  : 15/06/2026  │
     │  DUE DATE  : 15/12/2026  │
     │  STATUS  : ✓ IN CAL      │
     │  CAL BY   : Fluke Labs   │
     └──────────────────────────┘
  • GREEN label = In calibration / within due date.
  • RED label = Out of calibration / past due date. Equipment must NOT be used.
  • YELLOW label = Calibration due within 30 days (early warning).

3. Master Standards List

3.1 Master Standards with NIST / UKAS Traceability

Standard ID Type Description Range Traceability Calibration Frequency Custodian
MS-001 Reference Thermometer Fluke 1523 w/ PRT probe -50 to 300 °C NIST traceable (UKAS certificate) Annual (external) QA Lab
MS-002 Temperature Calibrator Dry-block calibrator (-20 to 150 °C) -20 to 150 °C NIST traceable Annual (external) Maintenance
MS-003 Mass Standards Set NIST Class 1 (1 mg – 1 kg) 1 mg – 1 kg NIST traceable Biennial (external) calibration QA Lab
MS-004 Mass Standards Set NIST Class F (1 kg – 20 kg × 5) 1–20 kg (×5) NIST traceable Biennial (external) QA Lab
MS-005 Pressure Calibrator Druck DPI 611 0–400 bar UKAS traceable Annual (external) Maintenance
MS-006 pH Buffers NIST-traceable buffer solutions (pH 4.0, 7.0, 10.0) pH 4.0, 7.0, 10.0 NIST SRM Monthly (replace solution) QA Lab
MS-007 Conductivity Standard KCl solution (1413 µS/cm ±0.5%) 1413 µS/cm NIST traceable Annual (replace) QA Lab
MS-008 Metal Detector Test Sticks Fe 1.0 mm, NF 1.5 mm, SS 2.0 mm Per stated dimensions Certified by manufacturer Annual (replace) QA Lab
MS-009 X-ray Test Piece SS spheres 0.5 mm, 1.0 mm Per stated dimensions Certified by manufacturer Annual (replace) QA Lab
MS-010 ATP Calibration Standard Manufacturer-supplied calibration kit Manufacturer certificate Annual (replace with kit) QA Lab
MS-011 Neutral Density Filters OD 0.5, 1.0, 1.5 Absorbance NIST traceable Biennial QA Lab
MS-012 Certified Stopwatch Quartz chronometer traceable to NIST 0–24 h NIST traceable Annual QA Lab

3.2 Master Standard Usage Rules

  1. Master standards are only used for calibration of working instruments — not for routine measurements.
  2. Master standards are stored in a controlled environment in the QA Lab (20 ± 2 °C, 45 ± 5% RH).
  3. Handling requires clean gloves. Standards kept in protective cases when not in use.
  4. Any damage or suspected damage to a master standard renders it unusable until re-certification.
  5. Log of master standard usage maintained in Master Standard Log (FS-CAL-F01).

4. Calibration Failure Procedure

4.1 Definition

A calibration failure occurs when a calibrated instrument, upon routine calibration, is found to be outside its required accuracy tolerance.

4.2 Immediate Actions

Step Action Responsible Timeline
1 Tag equipment OUT OF CALIBRATION (red label) and remove from service Calibration Technician Immediately
2 Identify all products, batches, or parameters measured by this instrument since its last valid calibration QA Manager Within 2 hours
3 Quarantine affected products / suspend affected process QA Manager Within 2 hours
4 Conduct out-of-tolerance investigation (Section 5) QA + Maintenance Within 48 hours
5 Adjust / repair / replace the instrument Maintenance / External Lab As soon as possible
6 Re-calibrate and return to service with green label Calibration Technician After repair
7 Release or disposition quarantined product based on investigation Food Safety Team Leader After investigation

4.3 Quarantine and Product Impact Assessment

The following information must be gathered to assess product impact:

Data Point Source
Date range of potentially affected measurements Calibration records (last valid cal date → today)
Product lots produced in that date range Production batch records
Critical limits affected (e.g., temperature, weight, metal detection) HACCP plan / process specs
Whether product was already released to customers ERP / despatch records
Customer notification needed? (if product may be unsafe) Food Safety Team decision

4.4 Escalation

If the calibration failure involves a CCP monitoring device (e.g., metal detector, temperature probe in a kill step), escalation to the Food Safety Team is mandatory within 2 hours. Regulatory notification may be required if affected product has been released.


5. Out-of-Tolerance (OOT) Investigation

5.1 Investigation Trigger

An OOT investigation is triggered when: - Calibration result exceeds the specified tolerance. - Calibration failure is identified during intermediate / daily check (e.g., checkweigher fails daily verification). - Equipment is found damaged, malfunctioning, or with expired calibration.

5.2 Investigation Form (FS-CAL-F02)

Section A — Incident Details | Field | Value | |---|---| | Equipment ID | | | Equipment Description | | | Parameter measured | | | Tolerance required | | | Calibration result (as found) | | | Date of last valid calibration | | | Date range possibly affected | |

Section B — Root Cause Analysis

Use one of the following methods: - 5 Whys: Iterative questioning to find root cause. - Fishbone (Ishikawa): Categories: Man, Machine, Method, Material, Measurement, Environment.

Root Cause Factor Finding
Operator error / misuse
Mechanical drift / wear
Environmental conditions (temp, humidity, vibration)
Damage (impact, moisture)
Inadequate calibration frequency
Software / firmware issue

Section C — Product Impact Assessment

Question Answer Evidence
Were any products measured during the OOT period? Y/N List batches
Are any critical limits compromised? Y/N Compare readings to CLs
Can product risk be quantified? Y/N Re-test data
Is any product already shipped to customers? Y/N Customer list
Is customer notification required? Y/N Regulatory assessment

Section D — Corrective & Preventive Actions (CAPA)

Action Owner Target Date Status
Repair / adjust instrument
Re-calibrate
Re-train operator(s)
Adjust calibration frequency
Add environmental protection
Other

Section E — Closure | Role | Name | Signature | Date | |---|---|---|---| | Investigation Lead | | | | | QA Manager Review | | | | | Food Safety Team Leader Approval | | | |

5.3 Calibration Frequency Adjustment

Based on OOT investigation findings, the calibration frequency may be adjusted: - If drift is minimal and stable: Frequency may be extended (e.g., from 6 months to 12 months) — requires > 2 years of stable data. - If OOT found: Frequency may be reduced (e.g., from annual to 6 monthly) until stability is demonstrated. - All frequency changes must be approved by QA Manager and documented on the calibration register.

Calibration results are trended annually to identify: - Drift direction / magnitude per instrument type. - Instruments approaching tolerance limits (predictive calibration). - Supplier / brand reliability differences. - Opportunities to extend or need to reduce calibration intervals.


Appendices

  • Appendix A: Calibration Procedure — Temperature (CAL-PROC-01)
  • Appendix B: Calibration Procedure — Thermocouple (CAL-PROC-02)
  • Appendix C: Calibration Procedure — Weights & Scales (CAL-PROC-03)
  • Appendix D: Calibration Procedure — Analytical Balance (CAL-PROC-04)
  • Appendix E: Calibration Procedure — Pressure (CAL-PROC-05)
  • Appendix F: Calibration Procedure — Steam Pressure (CAL-PROC-06)
  • Appendix G: Calibration Procedure — Flow Meters (CAL-PROC-07)
  • Appendix H: Calibration Procedure — pH (CAL-PROC-08)
  • Appendix I: Calibration Procedure — Conductivity (CAL-PROC-09)
  • Appendix J: Calibration Procedure — Metal Detector (CAL-PROC-10)
  • Appendix K: Calibration Procedure — X-ray (CAL-PROC-11)
  • Appendix L: Calibration Procedure — ATP Luminometer (CAL-PROC-12)
  • Appendix M: Calibration Procedure — ELISA Reader (CAL-PROC-13)
  • Appendix N: Calibration Procedure — Timers (CAL-PROC-14)
  • Appendix O: Calibration Procedure — Humidity (CAL-PROC-15)
  • Appendix P: Master Standard Log (FS-CAL-F01)
  • Appendix Q: Out-of-Tolerance Investigation Form (FS-CAL-F02)

End of Document — FS-CAL-001